FDA Recall Open, Classified

20 Gauge V-Lance Knife Model/Catalog Number: 8065912001 Software Version: N/A Product Description: Ophthalmic knife Component: N/A

Recall: Z-1922-2024 · Initiated April 22, 2024

Recall

Recall Number
Z-1922-2024
Event Number
94501
Firm
Alcon Research LLC Aspex Facility
FEI Number
1610287
Product Code
HNN
Status
Open, Classified
Root Cause
Process control
Initiated
April 22, 2024
Posted
May 30, 2024
Address
6201 South Fwy Fort, Worth, TX, 76134-2099

Description

20 Gauge V-Lance Knife Model/Catalog Number: 8065912001 Software Version: N/A Product Description: Ophthalmic knife Component: N/A

Reason

Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

Action

On April 22, 2024 VOLUNTARY MEDICAL DEVICE FIELD CORRECTION letters to customers. Actions to be taken by the Customer / User: If a knife does not perform as expected, or if you feel it does not meet your expectation for sharpness, discontinue use of that specific knife and replace the knife with a new one to avoid potential complications. To acknowledge your receipt of this Voluntary Medical Device Correction notification, please take the following steps: 1. Forward this notification to all departments or organizations using Alcon Ophthalmic Knives. 2. Follow the risk mitigation precautions provided in this notice when using identified catalogue numbers of ophthalmic knives. 3. Please complete the attached Response Form indicating your understanding of the included instructions and return the attached Response Form via email or fax to Alcon. Email: [email protected] Fax: 817-302-4337 Alcon has reported this issue to Health Authorities in accordance with applicable regulations. In the event you have experienced adverse events or product quality issues related to this communication, please contact Alcon via web (https://notifeye.alcon.com), by email ([email protected]) or by phone (1-800-757-9780). Adverse events or quality problems experienced with the use of this product may also be reported to the FDA MedWatch Adverse Event Reporting program either online (FDA EMAIL), by regular mail or by fax: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm. Should you have any questions or concerns about this matter or need help finding a replacement or substitute sterile standalone ophthalmic surgical knife, please call Alcon Customer Service or contact your Alcon Sales Representative.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom. Product distributed as sterile stand-alone knives have been placed on HOLD. Product included within Alcon Custom Pak (surgical procedure pack) are continuing to be distributed as ceasing distribution would cause immediate market shortage and result in canceled surgeries.

Quantity

29,190 units