FDA Recall Terminated

Innova 3100/ 3100IQ system with InnovaIQ Table Option Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures.

Recall: Z-1922-2010 · Initiated June 3, 2010

Recall

Recall Number
Z-1922-2010
Event Number
55552
Firm
Ge Healthcare, Llc
FEI Number
2126677
Product Code
MQB
Status
Terminated
Root Cause
Device Design
Initiated
June 3, 2010
Posted
June 29, 2010
Terminated
September 28, 2015
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Innova 3100/ 3100IQ system with InnovaIQ Table Option Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures.

Reason

GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.

Action

Consignees were sent on 6/3/10 a GE Healthcare "Urgent Medical device Correction" letter dated May 12, 2010. The letter was addressed to Hospital Administrators/ Risk Mangers, Managers of Radiology/Cardiology and Radiologists/ Cardiologists. The letter included the Safety Issue, Affected Product details, Safety Instructions, Product Correction and Contact Information. Customers are to contact a service representative immediately if they notice any cable damage.

Distribution

Worldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MI, MS, MO, MT, NE, NJ, NY, NC, OH, OR, PA, TN, TX, VT, WA, WV, and WI and countries of UNITED KINGDOM, SWEDEN, REPUBLIC OF KOREA, NORWAY, NEW ZEALAND, ITALY, ISRAEL, IRELAND, INDIA, HONG KONG, GERMANY, FRANCE, FINLAND, CZECH REPUBLIC, and AUSTRALIA.

Quantity

13