FDA Recall Terminated

Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of total cholesterol in serum and plasma.

Recall: Z-1921-2015 · Initiated May 7, 2015

Recall

Recall Number
Z-1921-2015
Event Number
71182
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
CHH
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 7, 2015
Posted
June 27, 2015
Terminated
March 24, 2017
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of total cholesterol in serum and plasma.

Reason

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Action

Roche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated labeling. Roche is not requesting return of the affected product. Customers may contact Todd Siesky Corporate Communications Director, Phone: 317.521.3911, Fax: 317.521.4696, E-mail: [email protected]. For technical questions, contact the Customer Support Center at 1-800-428-2336

Distribution

US (nationwide) including PR.

Quantity

6,255 analytical units distributed