FDA Recall Terminated

Automated External Defibrillator(AED) of brand name "Philips HeartStart HS1 OnSite" (Model number M5066A). Device is designed for use on a person in sudden cardiac arrest who is unresponsive when shaken and not breathing normally

Recall: Z-1916-2010 · Initiated April 30, 2010

Recall

Recall Number
Z-1916-2010
Event Number
55579
Firm
Philips Medical Systems
FEI Number
1000524572
Product Code
MKJ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 30, 2010
Posted
June 28, 2010
Terminated
December 10, 2012
Address
2301 5th Ave, Ste 200, Seattle, WA, 98121

Description

Automated External Defibrillator(AED) of brand name "Philips HeartStart HS1 OnSite" (Model number M5066A). Device is designed for use on a person in sudden cardiac arrest who is unresponsive when shaken and not breathing normally

Reason

Seventeen AEDs failed production line testing (Final Acceptance Test).

Action

The firm notified U.S. distributors on 4/30/2010 via phone call and e-mail to obtain consignee location information. A website providing information was activated on 4/30/2010. On 5/3/2010 Urgent Medical Device Recall letters, dated April 29, 2010 were sent to direct customers in the US. The letters explained the reason for the recall and Customers were asked to verify that they have the product subject to recall. Replacement AEDs will be provided at no charge to the customer. Field Safety Notices were attached to the letters and provided a list of actions for customers to take. Customers should check the serial number on their devices against the list provided, contact Philips Healthcare to arrange for replacement, and ensure that necessary staff are aware of the recall. For further information and support, Philips Healthcare should be contacted at 1-800-263-3342.

Distribution

Worldwide Distribution -- USA, Canada, Japan, United Arab Emirates, Austria, Switzerland, Germany, Spain, France, Italy, and Norway.

Quantity

205