FDA Recall Open, Classified

Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.

Recall: Z-1914-2025 · Initiated May 28, 2025

Recall

Recall Number
Z-1914-2025
Event Number
96927
Firm
Stryker GmbH Bohnackerweg 1 Selzach Switzerland
FEI Number
3002807830
Product Code
HSB
Status
Open, Classified
Root Cause
Process design
Initiated
May 28, 2025
Posted
June 6, 2025

Description

Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.

Reason

Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.

Action

Stryker notified consignees on about 05/28/2025 via FedEx letter. Consignees were instructed to review inventory and immediately quarantine any affected units on hand, and complete and return the Business Reply Form documenting receipt and product segregation. Upon receipt of the response form, Stryker will contact consignees to arrange for return of the affected units. If affected products have been further distributed, consignees were instructed to notify those customers and let Stryker know.

Distribution

Worldwide - US Nationwide distribution in the states of AL, FL, NJ, OR and the countries of Australia and Japan.