FDA Recall Terminated

Penner Manufacturing Transfer Electric Cascade with Scale, Patient Transfer/Lift System, Models 383000-1 (right entry) and 383000-1L (left entry). The firm name on the label is Penner Manufacturing Inc., Aurora, NE. For use with bathing systems in nursing homes, hospitals, and assisted living facilities to transfer or lift patients under the direct supervision of trained staff.

Recall: Z-1914-2011 · Initiated January 19, 2009

Recall

Recall Number
Z-1914-2011
Event Number
56977
Firm
Penner Mfg Inc
FEI Number
1922538
Product Code
FNG
Status
Terminated
Root Cause
Employee error
Initiated
January 19, 2009
Posted
April 8, 2011
Terminated
April 8, 2011
Address
102 Grant St, Ste A, Aurora, NE, 68818-3200

Description

Penner Manufacturing Transfer Electric Cascade with Scale, Patient Transfer/Lift System, Models 383000-1 (right entry) and 383000-1L (left entry). The firm name on the label is Penner Manufacturing Inc., Aurora, NE. For use with bathing systems in nursing homes, hospitals, and assisted living facilities to transfer or lift patients under the direct supervision of trained staff.

Reason

Incomplete weld on the pillar mounting bracket

Action

The firm, Penner, sent an undated letter flagged "Important Safety Inspection: Action Required" via certified mail on January 19, 2009 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect the Penner Patient Transfer for any missing Welds of the Pillar Mounting Bracket; inspect the areas indicated in the enclosed photos to confirm the bracket is welded on all edges to the main frame of the base; complete the inspection, and complete and return the attached WELD INSPECTION RESULTS OF PENNER PATIENT TRANSFER form via fax to 402 694-5844 or mail to Penner Manufacturing Inc, 102 Grant Street, Aurora, NE 68818, Attn: Quality Assurance Manager. Penner stated in the letter, if welds are not complete (no weld), discontinue use and contact the Quality Assurance Manager at Penner Manufacturing immediately at 402-694-5003. If you have any questions, contact the Quality Assurance Manager at 402-694-5003.

Distribution

Worldwide distribution: USA and countries including: Canada and Mexico.

Quantity

Model 383000-1 - 161 lifts; Model 383000-1L - 37 lifts