FDA Recall Terminated

Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of pediatric patients. The modules are color coded to accommodate patients of various sizes.

Recall: Z-1912-2009 · Initiated June 30, 2009

Recall

Recall Number
Z-1912-2009
Event Number
52694
Firm
Vital Signs Inc
FEI Number
2242551
Product Code
GAA
Status
Terminated
Root Cause
Device Design
Initiated
June 30, 2009
Posted
September 8, 2009
Terminated
August 27, 2012
Address
20 Campus Rd, Totowa, NJ, 07512

Description

Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of pediatric patients. The modules are color coded to accommodate patients of various sizes.

Reason

Not compatible: The luer connector on the Intraosseus needle provided does not fit securely on the luer connection of the IV Extension Set in the module.

Action

Vital Signs, Inc. issued an "Urgent Medical Device Correction" letter dated June 30, 2009 to all direct and indirect accounts by first class mail. Consignees were informed of the affected device, asked to complete a Confirmation form for return and to return unused product for replacement. For further questions, contact Vital Signs, Inc. at 1-973-790-1330 extension 404.

Distribution

Worldwide Distribution -- US, Canada, United Arab Emirate, Venezuela, Guam, Korea and United Kingdom.

Quantity

23,429 kits