Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V
Recall
- Recall Number
- Z-1911-2025
- Event Number
- 96727
- Firm
- Olympus Corporation of the Americas
- FEI Number
- 2429304
- Product Code
- NWB
- Status
- Open, Classified
- Root Cause
- Labeling design
- Initiated
- May 5, 2025
- Posted
- June 5, 2025
- Address
- 3500 Corporate Pkwy, Center Valley, PA, 18034-8229
Description
Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
Olympus notified consignees on 05/05/2025 via "URGENT Medical Device Correction" letter. Consignees were instructed to carefully review the notification, ensure all reprocessing and endoscopy unit personnel are completely knowledgeable and thoroughly trained on the updated reprocessing instructions, replace any copies of the reprocessing manual with the updated version (RC8415_02), complete and return the response form to recall web portal: https://olympusamerica.com/recall, enter the recall number: "0471" and complete the form as instructed, and forward the notification to affected customers if the affected product has been further distributed. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail [email protected].
Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.
12,195 units