FDA Recall Terminated

ARIA versions 8.0 and 8.1 Varian Medical Systems, Palo Alto, CA 94304 The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provided tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.

Recall: Z-1905-2011 · Initiated February 18, 2011

Recall

Recall Number
Z-1905-2011
Event Number
58112
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Other
Initiated
February 18, 2011
Posted
April 7, 2011
Terminated
June 19, 2012
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

ARIA versions 8.0 and 8.1 Varian Medical Systems, Palo Alto, CA 94304 The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provided tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.

Reason

An anomaly was identified in the ARIA version 8.0 and 8.1, when an image sequence template is added to a field in RT chart and a CIAO (Completely Irradiated Aperture Opening) is calculated, the resulting CIAO display at 4DITC may be incorrect.

Action

Varian Medical Systems sent an Urgent Medical Device Correction letter dated February 18, 2011, to all affected customers via Federal Express.. The letter identified the product, the problem, and the action to be taken by the customer. Varian has made design changes to address the issue in ARIA Radiation Oncology versions 8.2 and later. A corrective action will be undertaken to upgrade all affected items. For questions regarding this recall call US and Canada 1-888-VARIAN5 (888-827-4265. Europe +41 41 749 8844

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Algeria, Australia, Azerbaijan, Bahrain, Bangladesh, Barius, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, HOng Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lebanon, Lithania, Malaysia, Maurtama, Morocco, Netherlands, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Reunion, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Yemen.

Quantity

645