FDA Recall Terminated

Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237. Facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table.

Recall: Z-1904-2010 · Initiated June 14, 2010

Recall

Recall Number
Z-1904-2010
Event Number
55953
Firm
Mallinckrodt Inc
FEI Number
1518293
Product Code
IZO
Status
Terminated
Root Cause
Process control
Initiated
June 14, 2010
Posted
June 25, 2010
Terminated
February 25, 2012
Address
2111 E Galbraith Rd, Cincinnati, OH, 45237-1624

Description

Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237. Facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table.

Reason

Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative.

Action

Covidien Issued an "Urgent Device Correction" notification dated June 14, 2010 informing their customers of the potential problem. Consignees were instructed that Field Service Engineers will be in contact arrange a service visit to perform corrections on the affected units. For further information, contact Covidien Product Monitoring at 1-800-778-7898.

Distribution

Worldwide Distribution -- United States, Canada and China.

Quantity

144 units