Siemens Artis Q.zen ceiling-Interventional Fluoroscopic X-Ray System Model Number: 10848354
Recall
- Recall Number
- Z-1903-2019
- Event Number
- 83082
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 6, 2019
- Terminated
- February 9, 2021
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Siemens Artis Q.zen ceiling-Interventional Fluoroscopic X-Ray System Model Number: 10848354
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced.
Nationwide
15