FDA Recall Open, Classified

Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420 (b) DC Adapter, Product numbers: 1435, 1440 (c) AC Adapter, Product numbers: 1425, 1430 (d) Battery Pack, Product number: 1650DE Indication for Use for OUS HVAD System: The HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter.

Recall: Z-1903-2018 · Initiated May 2, 2018

Recall

Recall Number
Z-1903-2018
Event Number
80056
Firm
Heartware
FEI Number
3007042319
Product Code
DSQ
Status
Open, Classified
Root Cause
Device Design
Initiated
May 2, 2018
Posted
May 21, 2018
Address
14400 NW 60th Ave, Miami Lakes, FL, 33014-2807

Description

Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420 (b) DC Adapter, Product numbers: 1435, 1440 (c) AC Adapter, Product numbers: 1425, 1430 (d) Battery Pack, Product number: 1650DE Indication for Use for OUS HVAD System: The HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter.

Reason

Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.

Action

The firm notified physicians and healthcare professionals about the device correction by letter on 05/02/2018. The letter explained problem and provided recommendations for effective power source management of the system. Medtronic began notifying VAD Clinicians in the US on May 31, 2018, mainly via 2-day courier, and OUS via locally approved methods, of a newly available field service. A field service for application of a lubricant to the HVAD System power source connectors to mitigate the potential for transient interruptions of the electrical connection between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the HVAD Controller is now available.

Distribution

Worldwide Distribution including US Nationwide, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea, , Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, and Vietnam.

Quantity

175878 devices