FDA Recall Terminated

LIFEPAK 20 defibrillator/monitor. Product Part numbers: 3202487-xxx, 3202488-xxx, U3202487-000, 3202488-015, and U3202488-000.

Recall: Z-1903-2010 · Initiated May 27, 2010

Recall

Recall Number
Z-1903-2010
Event Number
55853
Firm
Physio Control, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
May 27, 2010
Posted
July 1, 2010
Terminated
October 21, 2014
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK 20 defibrillator/monitor. Product Part numbers: 3202487-xxx, 3202488-xxx, U3202487-000, 3202488-015, and U3202488-000.

Reason

A failure on the power supply assembly can result in either "No DC power" or "No DC or AC power." A failure of DC (battery) power can result in a delay of defibrillation therapy if no AC (line) power is available and the device will not operate.

Action

Firm began mailing Urgent Medical Device Correction notifications to affected customers on May 26, 2010 regarding the power supply to the device. All affected power supplies will be updated; this will be accomplished in two phases. Devices at a higher risk will be updated first in phase 1. The remaining lower risk devices will be updated afterward in Phase 2. Customers are advised to keep the defibrillators in service, and to follow recommended daily Operator Checklist steps, while service updates are scheduled. Customers with questions should contact the firm at 800-442-1142 between 6 am and 4 pm Pacific Time, Monday - Friday

Distribution

Worldwide distribution. In US, distributed domestically including to gov't accounts; and internationally to the following countries: ANGUILLA, ARGENTINA, AUSTRALIA, BAHAMAS, BAHRAIN, BARBADOS, BOLIVIA, BRAZIL, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CYPRUS, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, GERMANY, GUAM, GUATEMALA, HONG KONG, INDIA, IRAN, ISRAEL, JAMAICA, JORDAN, KUWAIT, LEBANON, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PUERTO RICO, SAUDI ARABIA, SOUTH AFRICA, SYRIA, TRINIDAD, TRINIDAD AND TOBAGO, TURKEY, UKRAINE, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS, and YEMEN.

Quantity

43,001 units