FDA Recall Open, Classified

Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900-001; V+C+Pro/Japan/PRT-00900-100; V+C/English/PRT-01098-000; V+O+C+S+N/English/PRT-01100-000; V+O+C+S+N/English/PRT-01100-001;

Recall: Z-1902-2021 · Initiated May 10, 2021

Recall

Recall Number
Z-1902-2021
Event Number
87838
Firm
Ventec Life Systems, Inc.
FEI Number
3013095415
Product Code
CBK
Status
Open, Classified
Root Cause
Device Design
Initiated
May 10, 2021
Posted
June 21, 2021
Address
22002 26th Ave Se, Bothell, WA, 98021-4903

Description

Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900-001; V+C+Pro/Japan/PRT-00900-100; V+C/English/PRT-01098-000; V+O+C+S+N/English/PRT-01100-000; V+O+C+S+N/English/PRT-01100-001;

Reason

Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.

Action

During the week of May 10, 2021, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter via FedEx or overnight mail to customers informing them that due to malfunctions of the Control Board there is the potential of unexpected shut downs and/or lost of ventilator blower function. As a result, the ventilators may fail to provide ventilation therapy to the patients. Customers are instructed to complete the following clinical action until a replacement unit can be provided: Follow your procedures to transfer the patient to an alternative source of ventilation, if available. If no other ventilation method is available, you may continue to use the affected product adhering to the guidelines below: 1. Use only when the patient is actively monitored to ensure timely response to a potential malfunction. 2. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 3. Use an external SpO2 monitor and set the alarm thresholds appropriately if the patient is oxygen dependent. 4. Connect the ventilator to an external monitoring system, wherever possible. In addition, Customers are to acknowledge receipt of the Customer Notification Letter by: 1. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to Ventec within 7 days. 2. Upon receipt of the Customer Response Form, Ventec will arrange shipment of replacement device(s) or arrange repair of existing device(s). 3. Return the affected product to Ventec per instructions and shipping label provided with replacement device(s). Contact Ventec Customer Support at 844-MYVOCSN (844-698-6276) to discuss alternative arrangements, if necessary. If customers are no longer the owner of any of these devices, they need to complete an ownership transfer form at VentecLife.com/tracking so that the Recalling Firm can follow up with the current owner of the impacted device(s). If customers have any question

Distribution

Distribution US nationwide, Hong Kong, and Japan

Quantity

1,934 devices (415 device that have not been updated)