FDA Recall Terminated

Dressing M/U XL, P/N 02344 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these products may have been sent with unsealed pouches by the contract manufacturer. Customers will be contacted and instructed to inspect their inventory for any product with unsealed pouches and discard them.

Recall: Z-1902-2017 · Initiated March 30, 2017

Recall

Recall Number
Z-1902-2017
Event Number
76994
Firm
Breg Inc
FEI Number
1000125378
Product Code
KGX
Status
Terminated
Root Cause
Employee error
Initiated
March 30, 2017
Posted
May 2, 2017
Terminated
June 23, 2017
Address
2885 Loker Ave E, Carlsbad, CA, 92010-6626

Description

Dressing M/U XL, P/N 02344 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these products may have been sent with unsealed pouches by the contract manufacturer. Customers will be contacted and instructed to inspect their inventory for any product with unsealed pouches and discard them.

Reason

They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.

Action

Breg sent an Urgent Medical Device Correction Letter to customers on 4/13/17. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: Please take the following actions: 1) Determine if any potentially affected products are in your stock by comparing each product's Lot and Date to the Lot Numbers and Expiration Dates in the table above. Refer to Picture 2 for the location of these fields on the product packaging. 2) Once you have matched Lot and Date, inspect the product to identify any open (unsealed) pouches. Refer to Picture 1 below. 3) Remove any open pouches from your inventory and discard. 4) Complete the Response Form included as Attachment A, and email or fax to Breg as follows: a. E-mail: [email protected]: Subject: Sterile Dressing Correction b. Fax: 800-FAX-BREG (329-2734)- Subject: Sterile Dressing Correction Customers with questions are instructed to contact Breg Customer Care at (800) 321-0607.

Distribution

US Nationwide Distribution

Quantity

2,609 units total