FDA Recall Terminated

Levitronix CentriMag Back-Up Console Article Number: 201-30200; extracorporeal blood circulation system.

Recall: Z-1902-2008 · Initiated March 17, 2008

Recall

Recall Number
Z-1902-2008
Event Number
47409
Firm
Levitronix, Llc
FEI Number
3003395347
Product Code
KFM
Status
Terminated
Root Cause
Device Design
Initiated
March 17, 2008
Posted
July 23, 2008
Terminated
July 6, 2010
Address
45 1st Ave, Waltham, MA, 02451-1105

Description

Levitronix CentriMag Back-Up Console Article Number: 201-30200; extracorporeal blood circulation system.

Reason

Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit

Action

Levitronix issued an Urgent - Voluntary Device Correction letter on 3/17/08 mailed to the US distributor and requested to contact their accounts. The letter revises the warning in the CentriMag Console Manual. Levitronix is issuing revised manuals. On March 13, 2008 the firm emailed and mailed (FedEx) to the one Canadian Hospital and Health Canada was informed of the Voluntary Advisory Notice via email on Wednesday, March 19, 2008. Rest of the World: Wednesday, March 12, 2008 through Monday, March 17, 2008. The Notice was issued as a Voluntary Advisory Notice after review and follow up with the UK MHRA. On 7/24/08 the firm issued a revised Recall Letter to advise users not to use Valley Lab Force FX-C or SSE2L Electrosurgery device with the firm's CentriMag Blood Pumping System and include the significant risk to health. The firm revised their letter and Black Box Warning and reissued the letters to the hospital, surgeons, etc. on 7/24/08. On 2/5/09 the firm issued a "Follow-Up to Urgent Device Correction of 3/17/08 " to advise users of the software upgrade to be installed in the CentrigMag Primary software Version 3.0 and CentriMag Back Up console , software version 2.0. Site visits will be made by Levitronix between Feb 2009 and April 2009 , software upgrade installed and new Operating Manuals provided.

Distribution

Worldwide Distribution: USA, Canada, Switzerland, UK, Italy, France, Greece, Spain, Slovenia, Czech, Turkey, South Africa, Argentina, Chile, Panama, Singapore, Taiwan, Israel, and Saudi Arabia.

Quantity

237 units