Siemens brand ACUSON S2000 Ultrasound System with software versions: VA10, VA10A, VA10B, VA10C, VA15; Model Number: 10041461; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Business Unit, Mountain View, CA The ACUSON SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications.
Recall
- Recall Number
- Z-1900-2011
- Event Number
- 51497
- Firm
- Siemens Medical Solutions USA, Inc.
- FEI Number
- 2936884
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Software change control
- Initiated
- January 14, 2009
- Posted
- April 6, 2011
- Terminated
- April 6, 2011
- Address
- 1230 Shorebird Way P.O. Box 7393, Mountain View, CA, 94043
Description
Siemens brand ACUSON S2000 Ultrasound System with software versions: VA10, VA10A, VA10B, VA10C, VA15; Model Number: 10041461; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Business Unit, Mountain View, CA The ACUSON SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications.
The product has a software problem in which previous patient measurement data gets associated with another patient's image.
The firm, SIEMENS, issued a "CUSTOMER SAFETY ADVISORY NOTIFICATION" letter to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to ensure that the first image in Exam Review is from the current patient; ensure that the Heart Cycle measurement tool is not selected on the Cale menu when entering Exam Review; be aware of this issue as described in the letter and perform four (4) or less measurements of the same measurement label. Should you have any questions, please call (650) 694-5398.
Worldwide distribution.