FDA Recall Terminated

GENESIS II TIBIAL DRILL GUIDE, REF 71440238, SIZE 5, QTY: (1), NON-STERILE, Smith & Nephew, Inc., Memphis, TN 38116, USA Drill guide used as manual surgical instrument during knee replacement surgeries.

Recall: Z-1900-2010 · Initiated April 20, 2010

Recall

Recall Number
Z-1900-2010
Event Number
55673
Firm
Smith & Nephew Inc
FEI Number
1020279
Product Code
LXH
Status
Terminated
Root Cause
Process control
Initiated
April 20, 2010
Posted
June 24, 2010
Terminated
December 7, 2011
Address
1450 E. Brooks Rd, Memphis, TN, 38116

Description

GENESIS II TIBIAL DRILL GUIDE, REF 71440238, SIZE 5, QTY: (1), NON-STERILE, Smith & Nephew, Inc., Memphis, TN 38116, USA Drill guide used as manual surgical instrument during knee replacement surgeries.

Reason

Markings for the drill guides are on the incorrect side: left is printed on right side and right is printed on the left side

Action

All affected Smith & Nephew consignees (International distributors) were notified of problem and the recall via email on 04/20/2010 and an Urgent: Medical Device Recall letter, dated 4/16/2010. The letter stated the reason for the recall and distributors were requested to immediately cease distribution and use of the affected products. They were also asked to examine their stock and stock of their accounts and quarantine any affected product. The Verification Section of the letter should be completed and returned even if there is no affected stock. Anyone the product was distributed to should also be notified and also return the product. Questions should be directed to Kelly Knight at 901-399-5441.

Distribution

International Distribution Only -- Belgium, Spain, Germany, Australia, and Brazil.

Quantity

12 units