PainSmart, PainSmart IOD Ambulatory Infusion System, Model Number: 360-1301
Recall
- Recall Number
- Z-1900-2008
- Event Number
- 48639
- Firm
- Curlin Medical Llc
- FEI Number
- 3002936294
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 30, 2008
- Posted
- September 22, 2008
- Terminated
- June 11, 2009
- Address
- 15751 Graham St, Huntingtn Bch, CA, 92649-1630
Description
PainSmart, PainSmart IOD Ambulatory Infusion System, Model Number: 360-1301
Over-delivery/free-flow involving Curlin ambulatory peristaltic multi-therapy infusion pumps where the pumping chamber door (platen) had become deformed in a manner that permitted gravity flow.
On May 30, 2008, Curlin Medical began distribution of a "Voluntary Field Correction Advisory Notice" to all customers in the United States via USPS Registered Mail. The Notice informed customers of the problem and gave the customers (1) Guidelines on testing pump volumetric accuracy, (2) Updated user manuals to elevate the "caution" statement related to dropped pumps to a "warning" statement, and (3) A warning label to be located on the pump: "Warning: Impact may cause damage. If dropped, pump must be checked for accuracy prior to use". A toll free number (888-287-5999) was given for customers experiencing issues that require immediate assistance.
Worldwide Distribution --- including USA and countries of Australia, Croatia, Germany, Italy, New Zealand, and South Africa.
39,366 units for all affected products/models total