FDA Recall Terminated

BD FlowSmart Set MiniMed Pro-set Rx Only Distributed by Medtronic MiniMed Inc. 8006464633 8185765555 Legal Manufacturer Becton, Dickinson and Company USA BD Medical-Diabetes Care Becton Dickinson France SAS

Recall: Z-1897-2017 · Initiated December 23, 2016

Recall

Recall Number
Z-1897-2017
Event Number
76294
Firm
Becton Dickinson & Company
FEI Number
2243072
Product Code
FPA
Status
Terminated
Root Cause
Device Design
Initiated
December 23, 2016
Terminated
November 9, 2017
Address
1 Becton Dr, Franklin Lakes, NJ, 07417-1815

Description

BD FlowSmart Set MiniMed Pro-set Rx Only Distributed by Medtronic MiniMed Inc. 8006464633 8185765555 Legal Manufacturer Becton, Dickinson and Company USA BD Medical-Diabetes Care Becton Dickinson France SAS

Reason

Patients are reporting high blood glucose (HBG) levels while using the MiniMed Pro-set and also noticed the catheter of the infusion set was bent/kinked.

Action

Becton Dickinson (BD) notified their sole customer Medtronic via e-mail on December 26, 2016. BD recommends that Medtronic notify their customers of the situation. For further questions, please call (201) 847-6800.

Distribution

US Distribution to the state of : CA

Quantity

57,750 units