FDA Recall
Terminated
BD FlowSmart Set MiniMed Pro-set Rx Only Distributed by Medtronic MiniMed Inc. 8006464633 8185765555 Legal Manufacturer Becton, Dickinson and Company USA BD Medical-Diabetes Care Becton Dickinson France SAS
Recall: Z-1897-2017
·
Initiated December 23, 2016
Recall
- Recall Number
- Z-1897-2017
- Event Number
- 76294
- Firm
- Becton Dickinson & Company
- FEI Number
- 2243072
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 23, 2016
- Terminated
- November 9, 2017
- Address
- 1 Becton Dr, Franklin Lakes, NJ, 07417-1815
Description
BD FlowSmart Set MiniMed Pro-set Rx Only Distributed by Medtronic MiniMed Inc. 8006464633 8185765555 Legal Manufacturer Becton, Dickinson and Company USA BD Medical-Diabetes Care Becton Dickinson France SAS
Reason
Patients are reporting high blood glucose (HBG) levels while using the MiniMed Pro-set and also noticed the catheter of the infusion set was bent/kinked.
Action
Becton Dickinson (BD) notified their sole customer Medtronic via e-mail on December 26, 2016. BD recommends that Medtronic notify their customers of the situation. For further questions, please call (201) 847-6800.
Distribution
US Distribution to the state of : CA
Quantity
57,750 units