FDA Recall Terminated

Boston Scientific Clipping Device, Resolution Clip, working length 235 cm, working channel 2.8 mm, Manufactured for Boston Scientific Corporation, Natick, MA; Product codes M00522610 (1 clip) and M00522612 (20 clips). The Resolution Clip Device is a pre-loaded single-use clip on a flexible delivery system. It is placed into the gastrointestinal tract for the purpose of endoscopic marking, hemostasis, anchoring jejunal tubes and closure of lumenal perforations.

Recall: Z-1896-2009 · Initiated June 17, 2009

Recall

Recall Number
Z-1896-2009
Event Number
52565
Firm
Boston Scientific Corporation
FEI Number
3001451463
Product Code
FHN
Status
Terminated
Root Cause
Process control
Initiated
June 17, 2009
Posted
September 2, 2009
Terminated
October 8, 2010
Address
500 Commander Shea Blvd, Quincy, MA, 02171-1518

Description

Boston Scientific Clipping Device, Resolution Clip, working length 235 cm, working channel 2.8 mm, Manufactured for Boston Scientific Corporation, Natick, MA; Product codes M00522610 (1 clip) and M00522612 (20 clips). The Resolution Clip Device is a pre-loaded single-use clip on a flexible delivery system. It is placed into the gastrointestinal tract for the purpose of endoscopic marking, hemostasis, anchoring jejunal tubes and closure of lumenal perforations.

Reason

The clip may not deploy or, if deployed, the clip may not release from the tissue.

Action

Consignees were notified by an "Urgent Recall Letter " dated June 17, 2009 and instructed to cease using the product because the calibration of equipment used to manufacture these products was not in control. This may potentially result in an inability to deploy or release the clip from the catheter. Consignees were instructed to return a business reply form for return authorization. Please contact Cindy Finney, Field Action Team Lead at 1-508-683-4678 for assistance.

Distribution

Nationwide, France, Germany, Great Britain, Italy, Netherlands, Spain and Sweden.

Quantity

4269