FDA Recall Terminated

PROBEAT-V, software versions 4.8.1S, and 5.1M Radiation therapy system.

Recall: Z-1892-2019 · Initiated April 15, 2019

Recall

Recall Number
Z-1892-2019
Event Number
83056
Firm
Hitachi America, Ltd., Power Systems Division
FEI Number
3003993895
Product Code
LHN
Status
Terminated
Root Cause
Software design
Initiated
April 15, 2019
Terminated
April 24, 2020
Address
1840 Old Spanish Trl, Houston, TX, 77054-2002

Description

PROBEAT-V, software versions 4.8.1S, and 5.1M Radiation therapy system.

Reason

There is a potential for a discrepant target position when using 3D3D matching mode in PIAS (Positioning Image Analysis System) software installed in the PROBEAT-V, proton therapy systems.

Action

The firm informed affected user sites of the recall beginning on April 15, 2019, via an Engineering Document. The firm distributed updated Urgent Medical Device Recall letters in June 2019. The firm stated that the PIAS software would be replaced with the corrected version. Until the software was corrected, users were asked to take the following interim countermeasures: -Limit the number of treatment planning CT slices to odd numbers only. - Create treatment plan only in Head First condition for new patients. - Use the beam limiting device in full open in the lateral direction (same as before). Customers were asked to complete and return the customer response form to the firm. The firm stated that the software correction will be implemented from June 17 - 29, 2019.

Distribution

Distributed to accounts located in AZ, MN, and TN. Foreign distribution to Japan.

Quantity

3