FDA Recall Terminated

EKOS EndoWave Infusion System consisting of a Drug Delivery Catheter and an Ultrasound Core. Labeled as Sterile, Single Use Only. The device is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

Recall: Z-1892-2008 · Initiated February 20, 2008

Recall

Recall Number
Z-1892-2008
Event Number
46923
Firm
EKOS Corporation
FEI Number
3001627457
Product Code
KRA
Status
Terminated
Root Cause
Process control
Initiated
February 20, 2008
Posted
September 16, 2008
Terminated
May 13, 2011
Address
11911 N Creek Pkwy S, Bothell, WA, 98011-8809

Description

EKOS EndoWave Infusion System consisting of a Drug Delivery Catheter and an Ultrasound Core. Labeled as Sterile, Single Use Only. The device is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

Reason

Product mislabeled - 6 cm treatment zone catheter had been built using 50 cm treatment zone drilled tubing. Consequently, drug was delivered outside of the treatment zone.

Action

On 02/20/08 firm notified customers by phone or email informing them of the issue and notifying them to segregate affected product and hold for return. Letters were sent via FedEx or email, or were hand delivered by firm representative to customers on 02/21/08 informing them of the issue and requesting return of affected product for replacement. Contact Jocelyn Kersten at 425-415-3132 if you have questions.

Distribution

Product was distributed to hospitals in the following states: WA, NC, IL, OH, NY, MI, and KY with no foreign distribution.

Quantity

10 units