FDA Recall Terminated

AVEA Standard ventilator, 17311-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.

Recall: Z-1889-2016 · Initiated May 17, 2016

Recall

Recall Number
Z-1889-2016
Event Number
74111
Firm
Carefusion 211 Inc dba Carefusion
FEI Number
3013421741
Product Code
CBK
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 17, 2016
Posted
June 13, 2016
Terminated
September 28, 2017
Address
22745 Savi Ranch Pkwy, Yorba Linda, CA, 92887-4668

Description

AVEA Standard ventilator, 17311-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.

Reason

CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).

Action

CareFusion sent a Field Safety Notice dated May 17, 2016, to all affected customers. The notice informed customers that CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM). This malfunction creates a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation to the patient. By design an audible clinical alarm is activated. Customers were informed of the problems identified and the actions to be taken. Customers were not required to return the affected devices or suspend use, customers would receive an copy of the identified affected serial numbers, as well as a response card, customers were requested to promptly return the response card to expedite the correction process and acknowledge receipt of the notification, Customers would be contacted by CareFusion to arrange onsite remediation of the affected devices, in the interim if any AVEA ventilator unit in their facility exhibits a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation, immediately remove the ventilator from service, provide alternate ventilation and contact CareFusion Technical Support per the contact information listed below to report the issue. Customers with questions should call 888-562-6016, for technical support 800-231-2466 or [email protected].

Distribution

Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.

Quantity

922 units total (501 units in US)