FDA Recall Terminated

M-Style Cup with Universal Vacuum Release (UVR) used with wall suction, Vacuum Assisted Delivery System (Ref: 10015LP). Product is intended to assist a clinician in the delivery of an infant during childbirth. Indicated for use during vaginal delivery and Cesarean sections.

Recall: Z-1888-2008 · Initiated January 16, 2008

Recall

Recall Number
Z-1888-2008
Event Number
46377
Firm
Cooper Surgical, Inc.
FEI Number
1216677
Product Code
HDB
Status
Terminated
Root Cause
Device Design
Initiated
January 16, 2008
Posted
September 17, 2008
Terminated
May 23, 2012
Address
95 Corporate Dr, Trumbull, CT, 06611-1350

Description

M-Style Cup with Universal Vacuum Release (UVR) used with wall suction, Vacuum Assisted Delivery System (Ref: 10015LP). Product is intended to assist a clinician in the delivery of an infant during childbirth. Indicated for use during vaginal delivery and Cesarean sections.

Reason

Vacuum Assisted Delivery System may have a loss of vacuum, or the cup will not release from the baby's head. A loss of vacuum may delay delivery and require the use of forceps and/or c-section for delivery. An inability to remove the cup from the head may delay delivery and/or prevent the user from applying forceps if necessary. This may result in serious injury or death to the baby.

Action

On 1/16/08 Cooper Surgical issued an Urgent Notice of Recall via Fed Ex for USA destinations and via e-mail and facsimile for international destinations, and firm confirmed receipt. The letter requests that accounts return recalled product. The recall was extended to three additional lots, and an Urgent Recall Notice dated 3/18/08 was issued. If you have any questions contact Thomas Williams at (203) 601-4741.

Distribution

Class II Recall - Worldwide Distribution --- including USA and Australia, Canada, Hong Kong, Latin America, Israel, and New Zealand.