FDA Recall Terminated

5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 75mm shaft length, 2B5ST

Recall: Z-1886-2018 · Initiated March 26, 2018

Recall

Recall Number
Z-1886-2018
Event Number
79801
Firm
Ethicon Endo-Surgery Inc
FEI Number
1527736
Product Code
GCJ
Status
Terminated
Root Cause
Process control
Initiated
March 26, 2018
Terminated
April 8, 2020
Address
4545 Creek Rd, Blue Ash, OH, 45242-2803

Description

5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 75mm shaft length, 2B5ST

Reason

Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.

Action

The firm notified customers and distributors by letter on March 26, 2018, and March 27, 2018, respectively. Customers were informed of the issue and potential risk of prolonged surgery or introduction of minute particles. Customers were instructed to: 1. Examine your inventory immediately to determine if you have any products subject to this recall on hand and quarantine such product(s). 2. Remove the products subject to this recall from your inventory and communicate the issue to all relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any of the devices subject to this recall have been forwarded to another facility, please contact that facility to arrange return. 3. Complete the Business Reply Form confirming receipt of this notice and fax or email it to Stericycle at 1-888-965-5794 or [email protected] within three (3) business days. Please return the BRF even if you do not have the product lot subject to this recall. 4. Customers are required to return all unused devices subject to this recall that are in their inventory immediately. 5. To return unused devices subject to this recall photocopy the completed BRF, place it in the box with the subject product(s), and affix the pre-paid authorized shipping label included with this recall notification letter. Distributors were instructed to: 1. Examine your inventory immediately to determine if you have the product lot subject to this recall on hand and quarantine such product(s). 2. Remove the product lot subject to this recall and communicate the issue to all relevant materials management personnel, or anyone else in your facility who needs to be informed. 3. Complete the Business Reply Form confirming receipt of this notice and fax or email it to Stericycle at 1-888-965-5794 or [email protected] within three (3) business days. Please return the BRF even if you do not have the product lot subject to this rec

Distribution

Nationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina.

Quantity

14873