Medtronic CB10W65R1, Custom Pack CB10W65R1 CH Access1/4, Custom Perfusion System, Sterile, Rx.
Recall
- Recall Number
- Z-1884-2021
- Event Number
- 87915
- Firm
- Medtronic Perfusion Systems
- FEI Number
- 1000116158
- Product Code
- DWF
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 29, 2021
- Terminated
- March 20, 2023
- Address
- 7611 Northland Dr N, Brooklyn Park, MN, 55428-1088
Description
Medtronic CB10W65R1, Custom Pack CB10W65R1 CH Access1/4, Custom Perfusion System, Sterile, Rx.
Specific lots may have been manufactured with insufficient or incomplete welds.
Medtronic notified customers by letters dated April 202, via 2-day UPS delivery to consignees explaining the reason for recall and requesting they immediately identify and quarantine all unused product in their inventory. The product was to be returned. A Customer Confirmation Certificate was enclosed for completion via scan and email. If product was further distributed or transferred, the consignee was requested to notify those downstream customers or facilities of the recall. For questions regarding this notification, customers were instructed to contact Smiths Medical via email at [email protected].
US Nationwide Distribution in the states of AL, AZ, CA, IN, KY, MA, OH, TN, and TX. There was no foreign/government/military distribution.
1 tubing pack