Dimension Vista System TSH Flex Reagent Cartridge (REF K6412). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin).
Recall
- Recall Number
- Z-1884-2013
- Event Number
- 65162
- Firm
- Siemens Healthcare Diagnostics, Inc
- FEI Number
- 1219913
- Product Code
- JLW
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 10, 2013
- Posted
- August 6, 2013
- Terminated
- May 10, 2016
- Address
- 333 Coney St, East Walpole, MA, 02032-1516
Description
Dimension Vista System TSH Flex Reagent Cartridge (REF K6412). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin).
A rare variant of TSH, identified in a small cluster of patients, is not detected.
On May 10, 2013, an Important Safety Information notice was sent to all affected Siemens Healthcare Diagnostic customers in the United States. On May 10, 2013, an Urgent Field Safety Notice was sent to all customers outside of the United States. These notices inform the customer that there is a rare variant of TSH, identified in a small cluster of patients that is not detected by some Siemens assays. This notice informs customers that they should continue to use the affected TSH assays as the observed rate of occurrence during a 30-month time period was 0.6 X 10 -7 and there are strong mitigating factors. These mitigating factors include co-interpretation of TSH results in conjunction with thyroxine and TS, patient history and clinical signs and symptoms. Although Siemens believes a look back is not required as the expected frequency of the variant is extremely rare, these notices state that this decision should be made by the medical director.
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
18584