FDA Recall Terminated

GE OEC 9800 Fluoroscopic X-Ray System, GE Healthcare, Surgery, Salt Lake City, Utah. The intended use of this device is to perform mobile fluoroscopic examination of human anatomy.

Recall: Z-1884-2008 · Initiated August 12, 2008

Recall

Recall Number
Z-1884-2008
Event Number
49486
Firm
OEC Medical Systems, Inc
FEI Number
1720753
Product Code
JAA
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
August 12, 2008
Posted
September 25, 2008
Terminated
April 23, 2012
Address
384 Wright Brothers Dr, Salt Lake City, UT, 84116-2862

Description

GE OEC 9800 Fluoroscopic X-Ray System, GE Healthcare, Surgery, Salt Lake City, Utah. The intended use of this device is to perform mobile fluoroscopic examination of human anatomy.

Reason

Beam limitation may be non-compliant on some X-ray units.

Action

Consignees were notified by letter on 08/12/2008. GE service representatives will check proper beam compliance on next scheduled maintenance visit or will schedule a visit to address the issue. Information available at 1-800-874-7378.

Distribution

Nationwide

Quantity

201 systems