FDA Recall
Terminated
Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.
Recall: Z-1883-2020
·
Initiated March 10, 2020
Recall
- Recall Number
- Z-1883-2020
- Event Number
- 85204
- Firm
- Northgate Technologies, Inc.
- FEI Number
- 1450997
- Product Code
- FFK
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 10, 2020
- Posted
- March 24, 2020
- Terminated
- February 2, 2021
- Address
- 1591 Scottsdale Ct, Elgin, IL, 60123-9361
Description
Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.
Reason
As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.
Action
Northgate Technologies, Inc. notified customers on 03/11/2020 via "Urgent: medical Device Recall Notification" email. The recall notice identified the affected and requested the customers to cease distribution and destroy affected product. The customers were asked to complete the "Recall Acknowledgment Form".
Distribution
US Nationwide distribution. No governmental.
Quantity
298 boxes. 3 units per box