FDA Recall Terminated

Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.

Recall: Z-1883-2020 · Initiated March 10, 2020

Recall

Recall Number
Z-1883-2020
Event Number
85204
Firm
Northgate Technologies, Inc.
FEI Number
1450997
Product Code
FFK
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 10, 2020
Posted
March 24, 2020
Terminated
February 2, 2021
Address
1591 Scottsdale Ct, Elgin, IL, 60123-9361

Description

Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.

Reason

As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.

Action

Northgate Technologies, Inc. notified customers on 03/11/2020 via "Urgent: medical Device Recall Notification" email. The recall notice identified the affected and requested the customers to cease distribution and destroy affected product. The customers were asked to complete the "Recall Acknowledgment Form".

Distribution

US Nationwide distribution. No governmental.

Quantity

298 boxes. 3 units per box