FDA Recall Terminated

GE Healthcare Definium 8000 Digital Radiographic System, model number 5135678. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 U.S.A. The intended use of this product is to generate tomographic images of human anatomy.

Recall: Z-1883-2008 · Initiated January 2, 2008

Recall

Recall Number
Z-1883-2008
Event Number
49487
Firm
GE Healthcare
FEI Number
2126677
Product Code
KPR
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
January 2, 2008
Posted
September 25, 2008
Terminated
December 10, 2011
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Definium 8000 Digital Radiographic System, model number 5135678. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 U.S.A. The intended use of this product is to generate tomographic images of human anatomy.

Reason

Not properly labeling the manual switch for disabling positive beam limitation (PBL) located on the overhead tube support user interface as required

Action

GE Field Engineer will visit each site to implement the correct label. Contact GE Healthcare at 1-262-544-3894 for assistance.

Distribution

AL, CA, CO, CT, DE, FL, GA, IA,, IL,, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI,WV and Washington D.C. OUS to include: Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Malta, Martinique (French), Netherlands, Norway, Palestine, Portugal, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, and Yugoslavia,

Quantity

668 (419 US / 249 OUS)