GE Healthcare Definium 8000 Digital Radiographic System.
Recall
- Recall Number
- Z-1882-2008
- Event Number
- 49503
- Firm
- GE Healthcare
- FEI Number
- 2126677
- Product Code
- KPR
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 7, 2008
- Posted
- September 25, 2008
- Terminated
- January 7, 2012
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Healthcare Definium 8000 Digital Radiographic System.
Radiation dose reading greater than actual dose: Error in the generator software on the GE Healthcare Revolution XR/d configured with HP 8200 PC consoles and all Definium 8000 systems may impact patient safety. Occasional generator software errors may cause light X-ray Images with reported mAs readings higher than was actually exposed to the patient using the overhead X-ray tube. The mAs rea
GE sent a Product Safety Notification dated March 7, 2008 to all customers. The letter described the safety issue, affected product, and stated that a GE Healthcare Field Engineer will schedule a field visit to install software updates.
Worldwide Distribution: USA, Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Martinique (French), Netherlands, Norway, Palestine, Portugal, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, and Yugoslavia.
454 (277 US / 177 OUS)