FDA Recall Terminated

GE Healthcare Definium 8000 Digital Radiographic System.

Recall: Z-1882-2008 · Initiated March 7, 2008

Recall

Recall Number
Z-1882-2008
Event Number
49503
Firm
GE Healthcare
FEI Number
2126677
Product Code
KPR
Status
Terminated
Root Cause
Software design
Initiated
March 7, 2008
Posted
September 25, 2008
Terminated
January 7, 2012
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Definium 8000 Digital Radiographic System.

Reason

Radiation dose reading greater than actual dose: Error in the generator software on the GE Healthcare Revolution XR/d configured with HP 8200 PC consoles and all Definium 8000 systems may impact patient safety. Occasional generator software errors may cause light X-ray Images with reported mAs readings higher than was actually exposed to the patient using the overhead X-ray tube. The mAs rea

Action

GE sent a Product Safety Notification dated March 7, 2008 to all customers. The letter described the safety issue, affected product, and stated that a GE Healthcare Field Engineer will schedule a field visit to install software updates.

Distribution

Worldwide Distribution: USA, Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Martinique (French), Netherlands, Norway, Palestine, Portugal, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, and Yugoslavia.

Quantity

454 (277 US / 177 OUS)