Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes TAVR PACK DYNJ69246B ABDOMINAL VASCULAR-LF DYNJ905291D ANGIOGRAPHIC PACK DYNJ86769 AORTAGRAM PACK DYNJ26783D AV FISTULA PACK CHS-LF DYNJ35615G AV SHUNT PACK-LF DYNJ0842516G CATHETER PACK DYNJ61926 EP DRAPE PACK-LF DYNJ47645C EP PACK DYNJ47782G EPS ACCESSORY PACK DYNJT2201S H CLINIC VASCULAR PK DYNJ33415A HEART FAILURE PACK DYNJ64877B HYBRID PACK DYNJ65925D INTERVENTIONAL RAD PACK DYNJ44100M INTERVENTIONAL RADIOLOGY PK-LF DYNJ0516014AG MV-PACEMAKER PACK-LF DYNJ41556F NEURO RADIOLOGY RICHMOND DYNJ68365D OR HYBRID-M DYNJ907758D, DYNJ907758F PACEMAKER DYNJ907960C PACEMAKER PACK DYNJ42557B DYNJ61411 DYNJ61411A DYNJ61411B DYNJ80160 PACER PACK DYNJ40952B PERIPHERAL ARTERIOGRAM RAD DYNJ61383B RADIOLOGY PACK DYNJ54762D TAVR PACK DYNJ55338B DYNJ69246A VASCULAR ACCESS TRAY DYNJ52606F VASCULAR TRAY DYNJ49802C VEIN PACK DYNJ57512 DYNJ58063 DYNJ84953
Recall
- Recall Number
- Z-1878-2026
- Event Number
- 98601
- FEI Number
- 1417592
- Product Code
- OEZ
- Status
- Open, Classified
- Root Cause
- Component change control
- Initiated
- February 27, 2026
- Posted
- April 17, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes TAVR PACK DYNJ69246B ABDOMINAL VASCULAR-LF DYNJ905291D ANGIOGRAPHIC PACK DYNJ86769 AORTAGRAM PACK DYNJ26783D AV FISTULA PACK CHS-LF DYNJ35615G AV SHUNT PACK-LF DYNJ0842516G CATHETER PACK DYNJ61926 EP DRAPE PACK-LF DYNJ47645C EP PACK DYNJ47782G EPS ACCESSORY PACK DYNJT2201S H CLINIC VASCULAR PK DYNJ33415A HEART FAILURE PACK DYNJ64877B HYBRID PACK DYNJ65925D INTERVENTIONAL RAD PACK DYNJ44100M INTERVENTIONAL RADIOLOGY PK-LF DYNJ0516014AG MV-PACEMAKER PACK-LF DYNJ41556F NEURO RADIOLOGY RICHMOND DYNJ68365D OR HYBRID-M DYNJ907758D, DYNJ907758F PACEMAKER DYNJ907960C PACEMAKER PACK DYNJ42557B DYNJ61411 DYNJ61411A DYNJ61411B DYNJ80160 PACER PACK DYNJ40952B PERIPHERAL ARTERIOGRAM RAD DYNJ61383B RADIOLOGY PACK DYNJ54762D TAVR PACK DYNJ55338B DYNJ69246A VASCULAR ACCESS TRAY DYNJ52606F VASCULAR TRAY DYNJ49802C VEIN PACK DYNJ57512 DYNJ58063 DYNJ84953
Unapproved design changes to the products outside of the 510(k) clearance.
On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
270,311 total