FDA Recall Terminated

Trident Endoscopic Ultrasonic Aspiration Needle (FNA)

Recall: Z-1878-2018 · Initiated March 12, 2018

Recall

Recall Number
Z-1878-2018
Event Number
79911
Firm
Micro-Tech (Nanjing) Co., Ltd. High Tech NO. 10 Gaoke Third Road Nanjing National Nanjing China
FEI Number
3004837686
Product Code
ODG
Status
Terminated
Root Cause
Device Design
Initiated
March 12, 2018
Terminated
August 30, 2018

Description

Trident Endoscopic Ultrasonic Aspiration Needle (FNA)

Reason

The combination of the firmness of the mass, the potential endoscope and elevator position combined with the force required to penetrate the mass may contribute to the breakage of the needle in the human body, which could cause injury.

Action

The firm emailed the distributor on March 12, 2018. (initial consignee). The consignee has been requested to return all products to the Manufacturer (Micro-Tech Nanjing).

Distribution

US distribution to one distributor in MI.

Quantity

40 total