FDA Recall Terminated

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.

Recall: Z-1877-2016 · Initiated April 5, 2016

Recall

Recall Number
Z-1877-2016
Event Number
73802
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
NJT
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 5, 2016
Terminated
November 30, 2017
Address
1700 Leider Ln, Buffalo Grove, IL, 60089-6622

Description

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.

Reason

Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.

Action

An Urgent: Medical Device Recall Notification dated 4/5/2016 was sent to consignees with the affected lots of CDX2 and RCC marker. The customers were notified of the problems as well as the following actions to be taken by the consignees: Do not use or continue to use the reagent lots listed above, as they have the potential to result in inadequate tissue staining and lead to weak or false negative pathology results. Appropriately destroy any unused or partially used affected lots of the reagent Complete the attached Recall Notification Acknowledgement Form that this action has been undertaken. The firm also request that the notification is passed to all those within their organization who need to be aware of this issue. The firm requests consignees to return the completed acknowledgement form to [email protected] to receive credit for affected lots that have been destroyed by the consignees laboratory.

Distribution

Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.

Quantity

90 Units