FDA Recall Terminated

ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Recall: Z-1876-2019 · Initiated May 10, 2019

Recall

Recall Number
Z-1876-2019
Event Number
82937
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
DSA
Status
Terminated
Root Cause
Process design
Initiated
May 10, 2019
Terminated
May 24, 2024
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Reason

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Action

The firm, GE Healthcare, initiated a "URGENT MEDICAL DEVICE CORRECTION"letter dated 05/10/2019. The letter explained the reason for recall and provided the following safety instructions: The affected ECG trunk cables and leadwires may continue to be used for monitoring only. Discontinue use of the affected ECG trunk cables and leadwires for patients where an arrhythmia that might require defibrillation is foreseeable, and in these patients use unaffected ECG trunk cables and leadwires. If defibrillation turns out to be unexpectedly needed when the affected ECG trunk cables and leadwires are being used just for monitoring, follow the instructions below: 1) Disconnect ALL ECG leadwires from the patient. 2) Defibrillate the patient per hospital protocol. 3) If it is possible to monitor the patients rhythm using defibrillator ECG leadwires or pads please do so. If this is not possible, reconnect ECG leadwires after the patients rhythm has been defibrillated. The recalled products will be replaced. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.

Quantity

58,045 units (31,108 US; 26937 OUS) in total