FDA Recall Terminated

Nipro Blood Tubing set with Priming Set and Transducer Protectors. Each device is packaged in a film pouch with 24 lines per case. The Nipro Set - Blood Tubing Set with Transducer Protector and Priming Set are disposable bloodlines intended to provide extracorporeal access to the patient s blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician in charge.

Recall: Z-1876-2018 · Initiated December 19, 2017

Recall

Recall Number
Z-1876-2018
Event Number
79834
Firm
Nipro Medical Corporation
FEI Number
1000132332
Product Code
FJK
Status
Terminated
Root Cause
Process control
Initiated
December 19, 2017
Terminated
October 19, 2021
Address
3150 NW 107th Ave, Doral, FL, 33172-2135

Description

Nipro Blood Tubing set with Priming Set and Transducer Protectors. Each device is packaged in a film pouch with 24 lines per case. The Nipro Set - Blood Tubing Set with Transducer Protector and Priming Set are disposable bloodlines intended to provide extracorporeal access to the patient s blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician in charge.

Reason

There is a possibility of the heparin line is occluded.

Action

Customers were notified on approximately December 19, 2017, via letter. Instructions included to place affected lots on hold and to advise Nipro if any affected product has been further distributed. If affected product has been further distributed, Nipro advised the customer to notify those customers and have the product returned immediately. Additionally Nipro requested that they complete and return the response form to arrange for product return. For further questions, please call (305) 599-7174.

Distribution

US Distribution to the states of : GA, NY, TN, TX.

Quantity

307,944 units.