Welch Allyn Vital Sign Monitor 5300 Series/Class II. (VSM 300). The VSM300 can monitor systolic and diastolic noninvasive blood pressure (NIBP), pulse rate, and mean arterial pressure (MAP). Units configured with the appropriate options can also simultaneously monitor temperature and noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and continuously monitor pulse rate. The monitor provides programmable audible and visual alarms and automatic NIBP measurements at selectable intervals.
Recall
- Recall Number
- Z-1873-2009
- Event Number
- 52721
- Firm
- Welch Allyn Protocol, Inc
- FEI Number
- 3023750
- Product Code
- MWI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 28, 2009
- Posted
- July 30, 2009
- Terminated
- August 10, 2011
- Address
- 8500 Sw Creekside Pl, Beaverton, OR, 97008
Description
Welch Allyn Vital Sign Monitor 5300 Series/Class II. (VSM 300). The VSM300 can monitor systolic and diastolic noninvasive blood pressure (NIBP), pulse rate, and mean arterial pressure (MAP). Units configured with the appropriate options can also simultaneously monitor temperature and noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and continuously monitor pulse rate. The monitor provides programmable audible and visual alarms and automatic NIBP measurements at selectable intervals.
The Welch Allyn VSM300 Vital Signs Monitor may experience a loss of audio which includes a loss of audio alarms.
An "Urgent: Correction" letter dated July 28, 2009 was sent to customers describing the risk associated with loss of audible alarms and recommend that the users mitigate the risk by checking for proper audible alarm function each time they power on the device. (Current labeling describes the variety of beeps and tones that should be present). Users were instructed: 1) if the audio fails, do not use the device, 2) Remove it from service and 3) contact Welch Allyn Protocol, Inc. at the following telephone number (1-800-289-2501) to arrange the return and repair of your device to Technical Services. Direct questions about this recall to Welch Allyn Protocol, Inc. by calling 1-503-530-7935.
Worldwide Distribution -- US, United Arab Emirates, Afghanistan, Austria, Australia, Belgium, Bulgaria, Bahrain, Botswana, Canada, Switzerland, China, Cyprus, Germany, Denmark, Egypt, Spain, Figi, France, Great Britain, Greece, Hong Kong, Croatia, Indonesia, Ireland, Israel, India, Iceland, Italy, Jordan, Kenya, Kyrgyzstan, South Korea, Kuwait, Lebanon, Sri Lanka, Malaysia, Nigeria, The Netherlands, Norway, New Zealand, Oman, Phillipines, Pakistan, Poland, Puerto Rico, Panama, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Thailand, Taiwan and South Africa.