FDA Recall
Open, Classified
PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile, Product Codes L923, L459, L493, L762, L463, L922, L495, L463, L497, L494, L493
Recall: Z-1872-2017
·
Initiated February 2, 2017
Recall
- Recall Number
- Z-1872-2017
- Event Number
- 76875
- Firm
- CP Medical Inc
- FEI Number
- 3012164473
- Product Code
- GAM
- Status
- Open, Classified
- Root Cause
- Unknown/Undetermined by firm
- Initiated
- February 2, 2017
- Address
- 1775 Corporate Dr, Ste 150, Norcross, GA, 30093-2960
Description
PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile, Product Codes L923, L459, L493, L762, L463, L922, L495, L463, L497, L494, L493
Reason
The tensile strength minimum as directed by USP <881> was not met over the length of the product's shelf life.
Action
The affected distributors/customers were contacted directly via telephone or by e-mail on February 6-7, 2017. Those end users not contacted by CP Medical directly will be contacted by their distributor
Distribution
AL, AZ, AR, CA, CO, CT, FL, ID, IL, IN, KY, LA, MT, NE, NJ, NM, NY, OK, OR, SC, TX, WA and South Africa
Quantity
504 boxes/12 sutures - 6,048 individual sutures