FDA Recall Terminated

Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.

Recall: Z-1872-2014 · Initiated May 19, 2014

Recall

Recall Number
Z-1872-2014
Event Number
68405
Firm
Stryker Endoscopy
FEI Number
2936485
Product Code
HRX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 19, 2014
Posted
June 25, 2014
Terminated
May 7, 2015
Address
5900 Optical Ct, San Jose, CA, 95138-1400

Description

Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.

Reason

STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.

Action

Device Correction letters were sent on May 23, 2014 to all users by certified mail. Internation sites will receive notification and be responsible for contacting customers.

Distribution

Worldwide Distribution.

Quantity

1992 trays