Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.
Recall: Z-1872-2014
·
Initiated May 19, 2014
Recall
- Recall Number
- Z-1872-2014
- Event Number
- 68405
- Firm
- Stryker Endoscopy
- FEI Number
- 2936485
- Product Code
- HRX
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 19, 2014
- Posted
- June 25, 2014
- Terminated
- May 7, 2015
- Address
- 5900 Optical Ct, San Jose, CA, 95138-1400
Description
Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.
Reason
STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
Action
Device Correction letters were sent on May 23, 2014 to all users by certified mail. Internation sites will receive notification and be responsible for contacting customers.
Distribution
Worldwide Distribution.
Quantity
1992 trays