FDA Recall Terminated

ACIST Automated Manifold Kit , REF Model BT2000, SKU # 014613, Sterile R, Made in Mexico, Manufacturer; ACIST LMedical Systems Inc, 7605 Fuller Road, Eden Prairie, MN .55344. The automated manifold kit is comprised of an injection manifold, a pressure transducer cartridge, a check valve, tubing (low, high and peristaltic pump), and a saline spike. This kit is designed for exclusive use with the ACIST CMS2000, E2000 Voyager and CVi models of Angiographic Contrast Delivery Systems. This kit is for single use.

Recall: Z-1872-2009 · Initiated May 11, 2009

Recall

Recall Number
Z-1872-2009
Event Number
52199
Firm
Acist Medical Systems
FEI Number
3001726808
Product Code
DXT
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 11, 2009
Posted
August 30, 2009
Terminated
February 4, 2012
Address
7905 Fuller Rd, Eden Prairie, MN, 55344

Description

ACIST Automated Manifold Kit , REF Model BT2000, SKU # 014613, Sterile R, Made in Mexico, Manufacturer; ACIST LMedical Systems Inc, 7605 Fuller Road, Eden Prairie, MN .55344. The automated manifold kit is comprised of an injection manifold, a pressure transducer cartridge, a check valve, tubing (low, high and peristaltic pump), and a saline spike. This kit is designed for exclusive use with the ACIST CMS2000, E2000 Voyager and CVi models of Angiographic Contrast Delivery Systems. This kit is for single use.

Reason

ACIST Medical Systems initiated a recall due to an increase in the incidence of field reports related to a bulge on the side and/or leaks in the bond of the 2.5 high-pressure tubing of BT2000 kits during a procedure which can result in unsatisfactory performance of the ACIST BT2000 Automated Manifold Kit.

Action

Consignees were sent on 5/11/09 a ACIST Medical Systems "Urgent: Medical Device Recall" letter dated May 11, 2009. The letter described the problem and the product. It advised consignees to discontinue the use or further distribution of the recall product and to return the inventory and the enclosed "Medical Device Recall Form". Please use the enclosed form with firms address: 7905 Fuller Rd, Eden Prairie, MD 55344. Additional questions please contact the firm at 1-877-272-2269, option 1.

Distribution

Worldwide Distribution including the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, HI, IL, ID, IN, IA, KS, KY, LA, MD, MI, MN, MS, MO, NE,NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT VA, WA, and WI.The countries of CANADA and HONG KONG

Quantity

62, 280 (57,110 US, 5130 Canada, 40 Hong Kong)