FDA Recall Terminated

Lactosorb Trauma Plating System, part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. The system is used in trauma and reconstructive surgical procedures in the mid face and craniofacial skeleton. Used to support non-load bearing tissues of cranial maxillofacial anatomy, including but not limited to brow fixation.

Recall: Z-1872-2008 · Initiated March 19, 2008

Recall

Recall Number
Z-1872-2008
Event Number
48461
Firm
Biomet Microfixation, Inc.
FEI Number
1032347
Product Code
JEY
Status
Terminated
Root Cause
Packaging process control
Initiated
March 19, 2008
Posted
September 16, 2008
Terminated
September 11, 2009
Address
1520 Tradeport Dr, Jacksonville, FL, 32218-2480

Description

Lactosorb Trauma Plating System, part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. The system is used in trauma and reconstructive surgical procedures in the mid face and craniofacial skeleton. Used to support non-load bearing tissues of cranial maxillofacial anatomy, including but not limited to brow fixation.

Reason

The outer foil pouch on the Lactosorb Plate system was not properly sealed. The product is labeled as sterile, but the lot was not sterilized. These devices are intended to be implanted into bone and are required to be sterile. A heavy bioburden would increase the potential for bone infection (osteomyelitis).

Action

On 3/19/08 the firm issued an Urgent Medical Device Recall Notice to its consignees by certified letters with return receipt. All returned products will be held in quarantine until final disposition is determined. If you have any questions contact Rachel Osbeck at 1-800-874-7711 or (904) 741-4400 ext 448.

Distribution

Class II Recall - Worldwide Distribution --- USA including states of NC, MI, TX, GA, HI, OK, IN and CA, and countries of Denmark, Germany, Mexico, Italy, Brazil, Japan, Canada, South Africa, China and the Philippines.

Quantity

117