FDA Recall Terminated

Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.

Recall: Z-1871-2013 · Initiated November 9, 2012

Recall

Recall Number
Z-1871-2013
Event Number
64985
Firm
SpineFrontier, Inc.
FEI Number
3005977257
Product Code
FZX
Status
Terminated
Root Cause
Device Design
Initiated
November 9, 2012
Posted
August 2, 2013
Terminated
June 5, 2015
Address
500 Cummings Ctr, Ste 3500, Beverly, MA, 01915-6516

Description

Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.

Reason

Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.

Action

Spine Frontier sent an Advisory Notice dated November 9, 2012, via email to all affected customers. Customers were advised that their FacetFuse Drill Guides need to be replaced. Customers were instructed to write their number ( included in the email) on the outside of the package and provide us with the tracking information. A replacement part has been sent out to you. If you have any further questions please call (978) 232-3990.

Distribution

US Distribution including Puerto Rico and the states of CO, KS, MA and TX.

Quantity

13 drill guides