FDA Recall Terminated

Accell Evo3C Demineralized Bone Matrix, 5 cc, Model Number: 02-6000-050 , Lot Number: 132160. Product Usage: Accell Evo3c Demineralized Bone Matrix Putty is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.

Recall: Z-1868-2014 · Initiated May 19, 2014

Recall

Recall Number
Z-1868-2014
Event Number
68330
Firm
Integra LifeSciences Corp.
FEI Number
3003418325
Product Code
MBP
Status
Terminated
Root Cause
Packaging
Initiated
May 19, 2014
Posted
June 25, 2014
Terminated
September 5, 2014
Address
311 Enterprise Dr, Plainsboro, NJ, 08536-3344

Description

Accell Evo3C Demineralized Bone Matrix, 5 cc, Model Number: 02-6000-050 , Lot Number: 132160. Product Usage: Accell Evo3c Demineralized Bone Matrix Putty is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.

Reason

This lot of Accell Evo3C Demineralized Bone Matrix, 5cc was incorrectly packaged in boxes that describe the product as Accell Evo3 Demineralized Bone Matrix. Evo3C is contraindicated for treatment of vertebral compression and not indicated for use in the posterolateral spine.

Action

Integra sent an Urgent Voluntary Medical Device Recall dated May 19, 2014 to customers via mail (traceable courier service). The recall notification letter identified affected product, reason for recall and actions to be taken Customers were instructed to complete the "RECALL ACKNOWLEDGEMENT AND RETURN FORM". and return to: [email protected] or FAX to 1-609-275-9445. For questions contact Customer Service at 1-800-550-7155.

Distribution

US Nationwide Distribution

Quantity

61