PHILIPS Xper Flex Cardio Physiomonitoring System, Model Numbers: 453564241901, 453564483321, 453564621791, and 989803199561 (international only) The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health care providers, such as physicians, registered nurses, cardiovascular technicians, radiology technicians, and other clinical health care providers.
Recall
- Recall Number
- Z-1867-2018
- Event Number
- 79611
- Firm
- Invivo Corporation
- FEI Number
- 1051786
- Product Code
- MWI
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 14, 2018
- Terminated
- April 30, 2021
- Address
- 12151 Research Pkwy, Ste 200, Orlando, FL, 32826-3286
Description
PHILIPS Xper Flex Cardio Physiomonitoring System, Model Numbers: 453564241901, 453564483321, 453564621791, and 989803199561 (international only) The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health care providers, such as physicians, registered nurses, cardiovascular technicians, radiology technicians, and other clinical health care providers.
The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.
The firm initiated their recall by letter on March 14, 2018. The Field Safety Notice informed the consignee about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients " the actions planned by Philips to correct the problem. If further information or support concerning this problem is needed, the consignee was directed contact the local Philips representative: 800-669-1328 option 2, then option 3.
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico., and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Italy, Kazakhstan, Kenya, Malaysia, Mexico, Mauritius, Netherlands, New Zealand, Pakistan, Philippines, Poland, Romania, Russian Fed, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom, Utd. Arab Emir, Uzbekistan and Vietnam.
4375 units