FDA Recall Terminated

Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone

Recall: Z-1862-2012 · Initiated May 17, 2012

Recall

Recall Number
Z-1862-2012
Event Number
61712
Firm
Stryker Medical Division of Stryker Corporation
FEI Number
1831750
Product Code
FNL
Status
Terminated
Root Cause
Component design/selection
Initiated
May 17, 2012
Posted
June 22, 2012
Terminated
May 12, 2014
Address
3800 East Centre Ave., Portage, MI, 49002

Description

Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone

Reason

An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac

Action

The firm, Stryker Medical, sent an "URGENT- Medical Device Correction" letter dated May 17, 2012 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to observe the beds for any malfunctions according to Preventive Maintenance Checklists in maintenance manual; report scale or bed exit malfunctions to Stryker (1-800-327-0770, option 2) to arrange for field service or to order service parts, at no charge. The customers were also instructed to forward this notice and advise Stryker of any subaccounts if the product was further distributed. A postage paid return postcard is to be filled for tracking and to make arrangements for repair. If you have any urgent questions or concerns, call 269-389-6466, Monday-Friday 8a.m-5p.m. (ET).

Distribution

Worldwide distribution:USA (nationwide) and countries including: Argentina, Bertec, Brasil, Brazil, Canada, Chile, China, India, Latin America, Mexico,Osteonics SA, Polska, Portugal, Puerto Rico, and Singapore.

Quantity

2546