Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Recall
- Recall Number
- Z-1862-2008
- Event Number
- 48164
- Firm
- Accuray Inc
- FEI Number
- 1000123585
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 1, 2008
- Posted
- September 16, 2008
- Terminated
- November 30, 2010
- Address
- 1310 Chesapeake Ter, Sunnyvale, CA, 94089-1100
Description
Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or [email protected] for additional information.
Product was distributed throughout the US and to China, Taiwan, Korea, Italy, Malaysia, Germany, Netherlands, Vietnam, Greece, Spain, and Puerto Rico.
105 units