FDA Recall Terminated

Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-1862-2008 · Initiated May 1, 2008

Recall

Recall Number
Z-1862-2008
Event Number
48164
Firm
Accuray Inc
FEI Number
1000123585
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
May 1, 2008
Posted
September 16, 2008
Terminated
November 30, 2010
Address
1310 Chesapeake Ter, Sunnyvale, CA, 94089-1100

Description

Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason

Couch may move unexpectedly, which may result in patient impacting the linear accelerator.

Action

All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or [email protected] for additional information.

Distribution

Product was distributed throughout the US and to China, Taiwan, Korea, Italy, Malaysia, Germany, Netherlands, Vietnam, Greece, Spain, and Puerto Rico.

Quantity

105 units