FDA Recall Terminated

Vertical Spring Arm for Single or Dual Monitor Mount (Harmony LL 500 and Harmony LL 700 surgical lighting systems), Steris Corporation. Model numbers YG19051 and YG19262.

Recall: Z-1850-2014 · Initiated April 4, 2014

Recall

Recall Number
Z-1850-2014
Event Number
68128
Firm
Steris Corporation
FEI Number
1527821
Product Code
FSY
Status
Terminated
Root Cause
Device Design
Initiated
April 4, 2014
Posted
June 23, 2014
Terminated
October 15, 2014
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

Vertical Spring Arm for Single or Dual Monitor Mount (Harmony LL 500 and Harmony LL 700 surgical lighting systems), Steris Corporation. Model numbers YG19051 and YG19262.

Reason

Potential for the welded seam in affected monitor spring arms to crack and/or break.

Action

On April 11, 2014 the firm sent Urgent Voluntary Field Correction Notices to their customers.

Distribution

Distribution to US (states: MA, NY, OH, PA & TX) and Canada.

Quantity

68 units