FDA Recall Open, Classified

Leica Surgical Operating Microscopes, Arveo, Part Number 10448999

Recall: Z-1849-2022 · Initiated August 26, 2022

Recall

Recall Number
Z-1849-2022
Event Number
90818
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
FSO
Status
Open, Classified
Root Cause
Component change control
Initiated
August 26, 2022
Posted
September 29, 2022
Address
10 Parkway North blvd, ste 300, deerfield, IL, 60089-6622

Description

Leica Surgical Operating Microscopes, Arveo, Part Number 10448999

Reason

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Action

The firm sent a Field Safety Notice to its consignees on 08/26/2022 by mail. The notice described the problem and stated that A Leica Microsystems representative will be contacting the consignee's to arrange this no-cost field update of the instrument. The field update will be scheduled to occur at your convenience within the next 6 months. The notice provided the following workaround: "Advice on action to be taken by the user: " You can continue to use the system pending this field update. Based on our analysis the probability is very remote that an unacceptable patient risk will occur due to the incorrect BrightCare Plus functionality caused by the non-conforming photodiode. " However, please ensure that all users of the affected microscope are aware of the Dangers of Use as described in section 3.4 and will follow the directions in section 5.3 ( Illumination ) and section 8 ( Operation ) of the user manual when handling a surgical operating microscope see excerpts below: o 3.4 Dangers of Use : &do not set the light intensity too high. & because if &light intensity is too high, uncontrolled tissue heating may occur. o 5.3 Illumination : &begin with a low light intensity and increase it until an optimum level of illumination is achieved. o 8 Operation : &use the lowest comfortable light intensity. The notice also requested acknowledgement of receipt of the Field Safety Notice as follows: "Please acknowledge receipt of this Field Safety Notice with the digital acknowledgment form by either scanning the QR code in your mobile phone or with the following link: https://forms.office.com/r/1WPXW93ugL."

Distribution

US Nationwide distribution.

Quantity

25 devices