FDA Recall Terminated

The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.

Recall: Z-1849-2015 · Initiated May 18, 2015

Recall

Recall Number
Z-1849-2015
Event Number
71451
Firm
Varian Medical Systems Inc
FEI Number
3011400544
Product Code
JAQ
Status
Terminated
Root Cause
Software Design Change
Initiated
May 18, 2015
Posted
June 26, 2015
Terminated
September 12, 2016
Address
501 Locust Ave, Charlottesville, VA, 22902-4869

Description

The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.

Reason

After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).

Action

All the consignees affected by this recall will be notified by: The Urgent Medical Device Correction Letter/ Field Safety Notification will be distributed to all affected users, with a description of the problem. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. Additionally, the letter will be posted to the customer support site: http:/ /www.MyVarian.com.

Distribution

Worldwide Distribution.

Quantity

55